• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER ALLCLEAR 2-DR DUO; STERRAD EQUIPMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED STERILIZATION PRODUCTS STERRAD 100NX STERILIZER ALLCLEAR 2-DR DUO; STERRAD EQUIPMENT Back to Search Results
Model Number 10104-008
Medical Device Problem Code Device Emits Odor (1425)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/25/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A field service engineer was dispatched to the customer site.However, details of the resolution are pending and asp will continue to follow up for the service resolution.Initial reporter phone #: (b)(6).Asp complaint ref #: (b)(4).
 
Event or Problem Description
A customer reported an event of an odor or smell emitting from the sterrad® 100nx sterilizer.There was no report of any injuries or human reactions.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Additional Manufacturer Narrative
A field service engineer was dispatched to the customer site.The vacuum tray assembly, vacuum control valve and oil return/solenoid valve were replaced to resolve the odor/smells issue.Unit meets specifications and was returned to service.H3: asp investigation summary: the investigation included a review of the device history record (dhr, complaint trending and system risk analysis (sra).The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Complaint trending for the issue of odor/smells was reviewed for the prior six months from open date and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for return and further evaluation.The assignable cause of the odor/smells is the vacuum tray assembly, vacuum control valve and oil return/solenoid valve.The field service engineer replaced these parts and confirmed the sterrad® 100nx was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD 100NX STERILIZER ALLCLEAR 2-DR DUO
Common Device Name
STERRAD EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA 92618
MDR Report Key9547485
Report Number2084725-2020-00003
Device Sequence Number2461238
Product Code MLR
UDI-Device Identifier10705037048700
UDI-Public10705037048700
Combination Product (Y/N)N
PMA/510(K) Number
K071385
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Remedial Action Notification
Type of Report Initial,Followup
Report Date (Section B) 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10104-008
Device Catalogue Number10104008
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2019
Supplement Date Received by Manufacturer02/13/2020
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date03/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1026-1027-2013
Patient Sequence Number1
-
-