MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PUNCTUA CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PUNCTUA CRT-D; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number N051
Medical Device Problem Codes Premature Discharge of Battery (1057); Device Displays Incorrect Message (2591)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 08/16/2019
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.No adverse patient effects were reported.This product remains in service.This report will be updated, should additional information be received.
 
Additional Manufacturer Narrative
The event date was updated, after device evaluation was completed.The returned implantable cardioverter defibrillator (icd) was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to this anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This device was not included in the advisory population.
 
Event or Problem Description
Additional information was reported indicating that this crt-d was removed and replaced.The icd was returned and a device evaluation was completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUNCTUA CRT-D
Common Device Name
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key9547716
Report Number2124215-2019-25458
Device Sequence Number19402014
Product Code LWP
UDI-Device Identifier00802526480928
UDI-Public00802526480928
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P010012/S255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/21/2013
Device Model NumberN051
Device Catalogue NumberN051
Device Lot Number100495
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Initial Date Received by Manufacturer 10/16/2019
Supplement Date Received by Manufacturer12/13/2019
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date02/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age46 YR
-
-