| Model Number |
N051 |
| Medical Device Problem Codes |
Device Displays Incorrect Message (2591); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/07/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient reported hearing tones from the cardiac resynchronization therapy defibrillator (crt-d).A remote monitoring interrogation found that the device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Device replacement was recommended.This device was explanted and returned for analysis.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This device was thoroughly inspected and analyzed upon receipt at our quality assurance laboratory.Review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.Malfunction of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to this anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This particular device was not included in the advisory population.
|
| |
|
Event or Problem Description
|
|
This report is being filed to submit the analysis results.
|
| |
|
Search Alerts/Recalls
|