MERGE HEALTHCARE MERGE CARDIO; PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS Back to Search Results
Model Number MERGE CARDIO V11.1.1
Medical Device Problem Code Data Problem (3196)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/05/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer's reported problem is currently under investigation by merge healthcare.The customer confirmed there were no allegations of patient harm due to this issue.This is still being investigated by support and development.When more information becomes available, a supplemental report will be submitted.
 
Event or Problem Description
Merge cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, hemodynamic studies and reports, measurements (via import from dicom structured reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.On (b)(6) 2019, a customer contacted merge healthcare and stated that diagnostic measurements (mr vena contracta and tr vena contracta) that were never taken on the modality were populating the cardio clinical report.Due to an incorrect values displaying in the diagnostic report, there is a potential for incorrect treatment of a patient that could result in harm.(b)(4).
 
Additional Manufacturer Narrative
This supplemental report is submitted to the fda in accord with applicable regulations and as indicated by merge healthcare in the initial report submitted (b)(6) 2020.Initial submission included that incorrect measurements and measurements that were never taken were population the cardio clinical report.Merge healthcare investigated this issue through a support request and post market product issue assessment.Investigation found the root cause of this issue to be due to misconfiguration of the affected systems caused by a change in the customer's environment where incoming measurements were mapped to the incorrect fields in the final report.An hhe was performed as there was no appropriate risk located in the risk document.The hhe found this issue to be an overall low risk for patient harm and the risk document was updated.The customer and support worked together to correct the measurement mapping strings which resulted in incorrect measurements on the clinical report.No further action is required.There were no reports of death, serious injury or injuries that were directly caused or contributed to as a result of this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERGE CARDIO
Common Device Name
PICTURE ARCHIVING AND COMMUNICATIONS SYSTEMS AND WORKSTATIONS
Manufacturer (Section D)
MERGE HEALTHCARE
900 walnut ridge drive
hartland WI 53029
MDR Report Key9548544
Report Number2183926-2019-00028
Device Sequence Number14024195
Product Code LLZ
UDI-Device Identifier00842000100751
UDI-Public00842000100751
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K051649
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date (Section B) 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMERGE CARDIO V11.1.1
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 12/05/2019
Supplement Date Received by Manufacturer05/07/2020
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date05/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-