• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH L-PLATE, LEFT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH L-PLATE, LEFT; PLATE, FIXATION, BONE Back to Search Results
Model Number 40-15091
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Injury (2348); Reaction (2414)
Date of Event 11/25/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device was not returned.If additional information becomes available, it will be provided in a supplemental report.Device disposition unknown.
 
Event or Problem Description
Open reduction and internal fixation of the left midfoot.Metallosis was observed during removal.
 
Additional Manufacturer Narrative
Correction: refer to h3 device discarded.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.In the case of metallosis, the opinion of a medical expert was requested.They stated the following; "[.] the literature talks about the varying corrosive effects of anodized titanium, relating to the thickness and color of the anodized layers.At a fracture site, wolff¿s law is occurring with a generation of an electrical current, although very slight, the current occurs within the solutions around the plate and soft-tissue and will corrode to metal surface.It is the interaction of an inorganic material in titanium plates, and soft tissue and the degree of roughness of the implant surface, the methods utilized in achieving the surface topography, and the chemistry of implant materials (titanium), [which] are some of the factors that may affect the soft tissue response of the plate to the adjacent soft tissue and create the grayish color.There has been no risk reported with this corrosive action since the levels of local metal ions and systemic ions is usually at very low levels in these cases.In case of high levels of metal ions, risks have been reported associated with local soft tissue damage, delayed hypersensitivity reaction (type iv) and systemic toxicity.[.]" metallosis cases can therefore be classified as patient related.They can be explained by the patient's reaction to the device.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device discarded.
 
Event or Problem Description
Open reduction and internal fixation of the left midfoot.Metalosis was observed during removal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
L-PLATE, LEFT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
CH  2545
MDR Report Key9549334
Report Number0008031020-2020-00087
Device Sequence Number2461208
Product Code HRS
UDI-Device Identifier07613154627998
UDI-Public07613154627998
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K063875
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number40-15091
Device Catalogue Number40-15091
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer03/09/2020
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date04/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age23 YR
-
-