|
It was reported that this right ventricular (rv) lead was surgically capped and replaced due to high thresholds and oversensing.And during the pre-operative check, it was determined that the pacemaker battery had gone from one year to end of life (eol) in six months.Subsequently, the device was also explanted and replaced.No adverse patient effects were reported.
|
|
Upon receipt at our post market quality assurance laboratory, detailed mechanical and electrical testing was performed on the device.The device battery status was ert.The device functioned normally throughout testing.Laboratory analysis determined that this device experienced normal battery depletion; however, the estimated longevity remaining value appeared to decrease more quickly than expected between routine follow-ups.Factors influencing the estimated longevity remaining calculation include pacing rate, amplitude, pulse-width and lead impedance.Any (even slight) changes in these factors will impact the battery consumption calculation and therefore the remaining longevity estimate.Please note that, despite the drop in estimated longevity remaining, the actual battery condition did not change significantly between follow-ups.In summary, it was determined that this device experienced normal battery depletion, but declared ert earlier than previously estimated due to fluctuating impedance measurements.
|