| Model Number |
S702 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
| Date of Event |
12/18/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
| |
|
Event or Problem Description
|
|
It was reported that this device exhibited premature battery depletion.A request was made to have data from this device analyzed, and a technical services (ts) consultant recommended device replacement.Subsequently, the patient underwent a revision procedure, and this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.A high current drain was detected.Detailed analysis confirmed that the high current drain was the result of a compromised capacitor, which resulted in the observed premature battery depletion.Investigation has determined that the high current drain is the result of compromised low voltage capacitors, which is caused by the presence of excess hydrogen from a liner component within the device.Boston scientific has issued an advisory communication regarding a subset of pacemakers in the accolade product family that has an elevated potential of exhibiting this behavior.This device is not part of the hydrogen induced premature depletion advisory population.
|
| |
|
Event or Problem Description
|
|
It was reported that this device exhibited premature battery depletion.A request was made to have data from this device analyzed, and a technical services (ts) consultant recommended device replacement.Subsequently, the patient underwent a revision procedure, and this device was explanted and replaced.No additional adverse patient effects were reported.Additional information: the device has been returned and analysis has been completed.This report has been updated with the product investigation results.
|
| |
|