| Catalog Number |
RONYX25026UX |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Myocardial Infarction (1969); Occlusion (1984)
|
| Date of Event |
02/28/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
During index procedure, 5 resolute onyx drug eluting stents were implanted; three in the rca and two in the lad.The following day, patient suffered elevated cardiac enzymes with no treatment carried out.Patient recovered.Cec adjudicated event as a non-q wave mi of the target vessel, 3rd udmi peri pci.Cec assessment commented that there was a large side branch occlusion.Investigator assessed the event is not related to device or anti platelet.Sponsor assessed event is possibly related to device but not related to anti platelet.
|
| |
|