COOK ENDOSCOPY D.A.S.H. DOMETIP DOUBLE LUMEN SPHINCTEROTOME; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
|
Back to Search Results |
|
Catalog Number DASH-35-480 |
Device Problem
Material Protrusion/Extrusion (2979)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/09/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation evaluation: a product evaluation was performed only by the picture provided with this report because the product said to be involved was not provided to cook for evaluation.The photo provided was in the endoscopic view.After a visual inspection of the device in the provided photo, the distal tip appears to be distorted/damaged.It is not known at what point the distal tip became distorted/damaged.Without return of the complaint device a complete evaluation of the product could not be performed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.Damage to the catheter tip can also occur if the precurved stylet is forcefully removed.The instructions for use instruct the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." if the elevator of the endoscope remains in the closed/up position when retraction of the sphincterotome is attempted and if additional pressure is applied, this could contribute to damage at the catheter tip.The instructions for use caution the user: "elevator should remain open/down when advancing or retracting sphincterotome." this activity will aid in device preservation.The instructions for use also advise the user to: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use." prior to distribution, all d.A.S.H.Dometip double lumen sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted and this represents an isolated occurrence.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
Event Description
|
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a cook d.A.S.H.Dometip double lumen sphincterotome.After the first cut, the physician found the top of the sphincterotome was peeling off.The procedure was completed with a new device.The photo provided depicts the distal tip of the sphincterotome protruding.A section of the device did not remain inside the patient's body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Search Alerts/Recalls
|
|
|