| Model Number |
VP-720-X |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Radiation Exposure, Unintended (3164); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
12/25/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic hepatectomy, when suturing the blood vessel, 2 hours into the procedure when both needle and thread broke.An x-ray was performed to locate the needle or confirmed that all pieces were located.The operation time was prolonged.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic hepatectomy, when suturing the blood vessel, 2 hours into the procedure, the needle was detached from thread.An x-ray was performed to locate the needle or confirmed that all pieces were located.Not all the pieces of the needle were retrieved.The operation time was prolonged for 10 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic hepatectomy, when suturing the blood vessel, 2 hours into the procedure, the needle was detached from thread.An x-ray was performed to locate the needle or confirmed that all pieces were located.There was no pieces fell into the patient's cavity.The operation time was prolonged for 10 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the sample noted one suture and two needles.One needle was disengaged from the suture.Upon microscopic inspection of the needle, normal crimp and swage marks were noted.As no sealed samples were returned, a needle attachment test could not be performed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|