| Model Number |
393009 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This device was explanted due to eri.High lv lead thresholds are believed to be the cause of battery depletion.No adverse patient events were reported.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.As of today, the medical device is not available for analysis, therefore the device itself could not be investigated.The information you provided has been entered into our quality system as a complaint.These types of complaints are used to evaluate systems and device performance throughout our organization and help to maintain and improve the performance of our devices.Should additional relevant information or the device itself become available, the investigation will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned and analyzed.The device memory demonstrated a normal device function while the device was implanted and in service.The bench test of the icd revealed that all therapy functions were available.The archive data and the programmed parameters were inspected.High left ventricular pacing thresholds, up to 7.5 v and 1.5 ms, were documented in the archive data.Therefore it is reasonable to assume that the pacing outputs were programmed accordingly.The last programmed left ventricular pacing output was 4.0 v and 1.5 ms.This represents a high current program, which results in a faster battery discharge.In conclusion the device was fully functional, there was no indication of a device malfunction.Analysis of the device revealed normal battery depletion.
|
| |
|
Search Alerts/Recalls
|