• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED-400-25
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Since the device was not returned, we are unable to perform further root cause analysis.All devices are 100% tested and all products are 100% inspected for damages and irregularities during manufacture.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received report that the pipeline flex stent was released completely, the marksman microcatheter retrieved the tip of guidewire, the tip of the guide wire was broken, the operator slowly withdrew the microcatheter and the guidewire together, and the stent and guide wire were removed.The pushwire was broken at the distal end.There was no friction or difficulty.No patient injury occurred.The devices were prepared and used per the instructions for use (ifu).This event occurred during the treatment of ophthalmic artery segment, unruptured aneurysm, that was saccular.The max diameter was 15mm and neck was 7mm.The distal landing zone was 3.6mm and the proximal was 4mm.The vessel anatomy was minimal in tortuosity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key9553044
Report Number2029214-2020-00005
Device Sequence Number11628492
Product Code OUT
UDI-Device Identifier00847536015968
UDI-Public00847536015968
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P100018.S011
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2022
Device Model NumberPED-400-25
Device Lot NumberA910396
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/03/2020
Initial Report FDA Received Date01/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age63 YR
-
-