MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Gas/Air Leak (2946)
Patient Problems Arrhythmia (1721); ST Segment Elevation (2059)
Event Date 12/12/2019
Event Type  Injury  
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed that at least seven applications were performed with balloon catheter, afapro28 with lot number 28147, on the date of the event.System notice 50012 ¿the refrigerant delivery path was obstructed¿ was triggered at the beginning of the ablation phase in the first application most likely due to residual moisture within cryoablation system.Eventually the moisture was cleared, and the subsequent applications were performed without any recorded anomalies.Clinical issues were encountered during procedure.In conclusion, the adverse events and alleged air ingress could not be assessed via data analysis.There is no indication of relation of adverse event to the performance and malfunction of the product.The physical product was not retuned for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, st elevation occurred, and the patient became bradycardic as the balloon catheter was inserted into the sheath.A small amount of air was aspirated.The patient was paced from the right ventricle and the st segments became normal immediately.The case was completed with cryo.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key9553270
MDR Text Key173773741
Report Number3002648230-2020-00015
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/26/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number83904
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2019
Initial Date FDA Received01/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/27/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
AFAPRO28 BALLOON CATHETER
Patient Outcome(s) Required Intervention;
-
-