| Model Number |
45-460 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
12/05/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade iv.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade iv.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: d.10., h.3., h.6.Device evaluation: visual analysis of the returned device identified: weight to the spec, crease fold, deformation, black particles, wear abrasion.Cloudy and bubbles were observed after autoclave disinfection process.Based on the device analysis the final assessment is: - no issues found related with the manufacturing process.
|
| |
|
Search Alerts/Recalls
|