ST. JUDE MEDICAL CATD SJM MASTERS SERIES VALSALVA AORTIC VALVED GRAFT; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23VAVGJ-515
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
As the physician was putting sutures through the sewing cuff of a 23vavgj product, sewing cuff began to pull away from valve.¿ it was like the sewing cuff was disintegrating¿.The valve was replaced with a 23 mm onx valved graft.
 
Event or Problem Description
As the physician was putting sutures through the sewing cuff of a 23vavgj product, sewing cuff began to pull away from valve.¿ it was like the sewing cuff was disintegrating¿.The valve was replaced with a 23 mm onx valved graft.
 
Additional Manufacturer Narrative
Additional information sections: b5, d10, h3, h6, h10.The reported event of the sewing cuff beginning to pull away from the valve was confirmed.The investigation found one of the loops of the suture holding the cuff in place was broken, which could cause loosening of the cuff.The knots of the suture were intact and no evidence was found to suggest that the damage occurred during manufacturing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALSALVA AORTIC VALVED GRAFT
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
MDR Report Key9553938
Report Number2648612-2020-00002
Device Sequence Number19322800
Product Code LWQ
UDI-Device Identifier05414734009522
UDI-Public05414734009522
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2023
Device Model Number23VAVGJ-515
Device Catalogue Number23VAVGJ-515
Device Lot Number6890040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/03/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/20/2019
Supplement Date Received by Manufacturer02/25/2020
Initial Report FDA Received Date01/06/2020
Supplement Report FDA Received Date03/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-