| Model Number |
23VAVGJ-515 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
As the physician was putting sutures through the sewing cuff of a 23vavgj product, sewing cuff began to pull away from valve.¿ it was like the sewing cuff was disintegrating¿.The valve was replaced with a 23 mm onx valved graft.
|
| |
|
Event or Problem Description
|
|
As the physician was putting sutures through the sewing cuff of a 23vavgj product, sewing cuff began to pull away from valve.¿ it was like the sewing cuff was disintegrating¿.The valve was replaced with a 23 mm onx valved graft.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information sections: b5, d10, h3, h6, h10.The reported event of the sewing cuff beginning to pull away from the valve was confirmed.The investigation found one of the loops of the suture holding the cuff in place was broken, which could cause loosening of the cuff.The knots of the suture were intact and no evidence was found to suggest that the damage occurred during manufacturing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.
|
| |
|