| Model Number |
NOT APPLICABLE |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
Missing Value Reason (3192)
|
| Date of Event |
12/07/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial report indicated that the skin port had separated from the a-tube.Subsequent email described the a-tube as separated at the bumper.There is no information to suggest that the a-tube which had been implanted for 20 months had any signs of material degradation.The skin port attachment to the a-tube and the a-tube overmold are both designed and tested to ensure that normal forces which might be applied to the device do not result in separation with an appropriate safety margin.Without having the device to evaluate, it is not possible to determine the root cause of this event.The patient underwent endoscopy to remove the device from the stomach.There were no issues with removal of the tube and the procedure was well tolerated.
|
| |
|
Event or Problem Description
|
|
Patient is (b)(6) female with a history of obesity undergoing aspire assist therapy, a-tube placement in (b)(6) of 2018.Rn reported being contacted by a patient who stated that the skin port had come off and the a-tube had retracted into the stomach.
|
| |
|