| Model Number |
1516-30-410 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Edema (1820); Pain (1994); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
|
| Date of Event |
05/08/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
| |
|
Event or Problem Description
|
|
The patient underwent a second revision of the right knee on (b)(6) 2019 secondary to pain, swelling and suspected loosening.Intraoperatively, the surgeon noted the patellar component was well-fixed; the posterior cruciate ligament was lax and there was significant placement of the tibia in relation to the femur and a 90-degree flexed position; and the tibial component was loose at the cement to implant interface.No intraoperative complications were noted.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # : (b)(4).Investigation summary: no device associated with this report was received for examination.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|