Model Number 254622
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(6).Udi: (b)(4).
 
Event or Problem Description
It was reported by the affiliate via phone that during an acl reconstruction when the biointrafix 6-7mm x 30mm tprscr was inserted, it was broken off.All the pieces were removed from the patient.The procedure was completed using another device.No patient consequence and no surgical delay was reported.No additional information was provided.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that the biointrafix 6-7mmx30mmtprscr during the surgery of acl reconstruction,when insert the device, the device was broken off.All the pieces were retrieved from the patient.The complaint device was received and inspected.Visual observation confirms the distal tips of both the screws are completely broken and there is a crack diagonally along the length of the screw.It revealed presence of tissue or bio debris on the device and its length indicating this device has been already used.We can be confirmed this complaint.The possible cause of this failure is consistent with the hard bone quality of the patient.We cannot discern a root cause outside this possibility.A manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot: a manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D10, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOINTRAFIX 6-7MMX30MMTPRSCR
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9555662
Report Number1221934-2020-00054
Device Sequence Number19698409
Product Code HWC
UDI-Device Identifier10886705013197
UDI-Public10886705013197
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K032167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2020
Device Model Number254622
Device Catalogue Number254622
Device Lot NumberL502582
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2020
Initial Date Received by Manufacturer 12/26/2019
Supplement Date Received by Manufacturer02/17/2020
02/18/2020
Initial Report FDA Received Date01/06/2020
Supplement Report FDA Received Date02/18/2020
02/19/2020
Patient Sequence Number1
Patient Age36 YR
Patient Weight80
-
-