Catalog Number H938724E
Device Problems Break (1069); Crack (1135)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was manufactured from july 17, 2018 ¿ july 19, 2018.The device was received for evaluation.Unaided visual inspection was performed which observe a damaged or cracked silicone outlet tubing.The reported condition was verified during initial inspection.The cause of the condition could not be determined.A functional testing was not performed for this complaint.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
During evaluation it was identified that a em2400 valve set had a damaged/cracked silicone outlet tubing.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUROPEAN VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9555898
MDR Text Key173938002
Report Number1416980-2020-00009
Device Sequence Number1
Product Code LHI
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 01/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724E
Device Lot Number60138424
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/21/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/06/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-