MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SELF CENT HIP 40X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SELF CENT HIP 40X28 RUST; HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI Back to Search Results
Model Number 1035-40-000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Anemia (1706); Edema (1820); Hematoma (1884); Pain (1994); Seroma (2069); Joint Dislocation (2374); No Code Available (3191)
Date of Event 12/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Patient received implant unknown hip side on (b)(6) 2019.Revision on (b)(6) 2019 because of pain and luxation.The pe ring was found to be fractured.Further details unknown at the moment.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  patient code: no code available (3191) used to capture device revision or replacement.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : following review of the information received, it was concluded that it was unlikely that a product issue as a result of the manufacture or design was present.The complaint shall be closed with an unjustified conclusion, it shall be entered onto the complaints database and monitored through trend analysis.Should further information be received, then the complaint shall be reviewed and further investigation completed as required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.  additional narrative:.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF CENT HIP 40X28 RUST
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR POLY/METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9557504
Report Number1818910-2020-00769
Device Sequence Number2275637
Product Code KWY
UDI-Device Identifier10603295003304
UDI-Public10603295003304
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K812672
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2024
Device Model Number1035-40-000
Device Catalogue Number103540000
Device Lot NumberJ4339Z
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2020
Initial Date Received by Manufacturer 12/13/2019
Supplement Date Received by Manufacturer02/03/2020
02/18/2020
02/18/2020
03/11/2020
Initial Report FDA Received Date01/07/2020
Supplement Report FDA Received Date02/11/2020
03/03/2020
03/09/2020
03/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 NEUT 40IDX60OD; ARTICUL/EZE BALL 28 +8.5 BL; ALTRX +4 NEUT 40IDX60OD; ARTICUL/EZE BALL 28 +8.5 BL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
-
-