• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. MINI TRAY STD COCR +3 OFFSET; PROSTHESIS, EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. MINI TRAY STD COCR +3 OFFSET; PROSTHESIS, EXTREMITIES Back to Search Results
Model Number N/A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994); No Information (3190)
Date of Event 12/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: item#113634; comp primary stem 14mm mini; lot# 300540, item# 110031424; cr vivacit-e 36mm brng std; lot# 64343846, item# 115310; comp rvrs shldr glnsp std 36mm; lot# 664970, item# 115395; comp rvs cntrl 6.5x25mm st/rst; lot# 139000, item# 180558; comp nlk scr 3.5hex 4.75x20 st; lot# 236640, item# 180552; comp lk scr 3.5hex 4.75x25 st; lot# 795520, item# 180551; comp lk scr 3.5hex 4.75x20 st; lot# 641130, item# 115330; comp rvrs shdr glen bsplt +ha; lot# 760180, item# 118001; versa-dial/comp ti std taper; lot# 349060.Report source: foreign: event happened in (b)(6).The product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2019-05379.Remains implanted.
 
Event or Problem Description
It was reported patient suffered pain due to disassociated prosthesis approximately one month post-implantation.Due to short notice, replacement implants and instruments were not available.The surgeon reattached the glenosphere and humeral bearing to baseplate, used 2 juggerknots to help stabilize the shoulder.Two cerclage wires were attached to site.Patient will be observed to see how he does.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Correction: implant date is corrected.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: a2; b4; b5; g4; g7; h1; h2; h3; h6 corrected: h4 no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified separation of the glenosphere from the base plate of the reverse-type right shoulder arthroplasty and the humerus is slightly elevated in relation to the glenoid.Dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TRAY STD COCR +3 OFFSET
Common Device Name
PROSTHESIS, EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9558399
Report Number0001822565-2019-05380
Device Sequence Number2578374
Product Code KWS
UDI-Device Identifier00887868231346
UDI-Public(01)00887868231346
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K181611
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110031402
Device Lot Number64327562
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/12/2019
Supplement Date Received by Manufacturer01/14/2020
06/04/2020
Initial Report FDA Received Date01/07/2020
Supplement Report FDA Received Date02/06/2020
06/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age72 YR
-
-