• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CLEARLINK CONTINU-FLO SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2C8519
Device Problem Filling Problem (1233)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported a paclitaxel set under infused due to a ¿kink below drip chamber when the medication and tubing were place inside bags used to transport pre-mixed product¿.The expected infusion time was thirty minutes; however, the actual infusion time was fifty minutes.The device was filled with 100ml of avastin and infused via a baxter pump.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information: correction made: it was reported a clearlink system continu-flo solution set (previously paclitaxel set) under infused due to a ¿kink below drip chamber when the medication and tubing were place inside bags used to transport pre-mixed product¿.The expected infusion time was thirty minutes; however, the actual infusion time was fifty minutes.The device was filled with 100ml of avastin and infused via a baxter pump.There was no report of patient injury or medical intervention associated with this event.No additional information is available.The device was not returned and the lot number is unknown; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK CONTINU-FLO SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - CARTAGO
see h10
cartago 30106
CS   30106
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9559766
MDR Text Key174108633
Report Number1416980-2019-07443
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961225
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2C8519
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/29/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-