There was no known patient involvement.
The heater-cooler 16-02-80 is not distributed in the usa, but it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601).
Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.
The z number is z-2076/2081-2015.
Livanova (b)(4) manufactures the heater-cooler system 3t.
The incident occurred in (b)(6).
Through follow-up communication livanova (b)(4) learned that the device was cleaned regularly per the instruction for use and that it was placed inside the operating theatre during use with the fan opposite to the sterile field.
The estimated distance between the device and the surgery field is about 3 meters.
A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.
If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|