Model Number 228152 |
Device Problem
Positioning Failure (1158)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/18/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
This is report 4 of 6 for the same event.It was reported by the affiliate in (b)(6) that during a meniscal repair 6 truespan 24 degree peek were not able to be deployed.The procedure was not able to be completed and the surgeon had to performed a menisectomy.There was no delay in the surgical procedure.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for tned that was not availab e ff tformation is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: b5: subsequent follow-up with the customer, additional information was received.It was reported that all devices were discarded, none were implanted.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: additional information: b5: subsequent follow-up with the customer, additional information was received.It was reported that there was no surgical delay.It was reported that there were no patient consequences or impact to the user or patient.It was reported that there was no surgical intervention planned (e.G.X-rays, additional/change in procedures, no prescriptions, otc, revisions).It was reported that the age and activity of the patient were unknown.It was reported that the patient had extensive complex tearing.It was reported that the patient consented for menisectomy if required.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.Possible root causes for the reported failure of not deploying could be over penetrating the needle causing the silicon tube to move, or not inserting the needle to the proper depth for deployment.However without the return of the complaint device, and no further information provided we cannot determine a definitive root cause.A manufacturing record evaluation was performed for the finished device lot number (5l23182), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
|