• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT20649-2
Device Problem Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/13/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that unexpected receiver shutdown occurred. No product or data was provided for evaluation. Confirmation of the allegation and a probable cause could not be determined. No injury or medical intervention was reported.
 
Manufacturer Narrative
(b)(4).
 
Event Description
The product was evaluated. An external visual inspection was performed and passed. Charge and boot tests were performed and failed due to the receiver not turning on, but will turn on with a known good battery. An internal visual inspection was performed and failed due to moisture damage. Pictures were taken. The receiver log was downloaded and reviewed finding error related to the complaint. The allegation was confirmed. The probable cause was determined to be a damaged battery. No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
Manufacturer Contact
ashley spoto
6340 sequence dr.
san diego, CA 92121
8582000200
MDR Report Key9564503
MDR Text Key174949041
Report Number3004753838-2020-002063
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Reporter Occupation
Type of Report Initial,Followup
Report Date 03/03/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberMT20649-2
Device Catalogue NumberSTR-GL-102
Device Lot Number5224499
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received02/07/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-