• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION BALLOON CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS GOLD PTA DILATATION BALLOON CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number ATG80224
Medical Device Problem Codes Deflation Problem (1149); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Needle Stick/Puncture (2462); Device Embedded In Tissue or Plaque (3165)
Date of Event 12/10/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 12/2021).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac vein, the pta balloon allegedly detached from the catheter and remained inflated.It was further reported that the catheter was pulled out but the balloon remained inside the patient.Reportedly the balloon had to be punctured with a needle in order to be deflated.The physician then used a 12fr sheath to snare the balloon out and complete the procedure.There was no reported complications to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD PTA DILATATION BALLOON CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9564606
Report Number2020394-2019-05910
Device Sequence Number19414685
Product Code LIT
UDI-Device Identifier00801741060977
UDI-Public(01)00801741060977
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG80224
Device Catalogue NumberATG80224
Device Lot Number93ZA0090
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2019
Initial Report FDA Received Date01/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
-
-