| Model Number |
199723640 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional device product code: mni.(b)(4).The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from synthes reports an event in (b)(6) as follows: it was reported that the stabilization system implanted on the patient was dismantled due to the breaking of two tightening nuts and it was necessary to re-operate the patient to remove and replace the faulty system.The devices were removed.No further information provided.Upon receipt of the devices at manufacturing site it was identified that 5.5 exp verse fen screw has torn threads.Concomitant devices: unknown set screw (part #: unknown, lot #: unknown, quantity: unknown) and unknown rod (part #: unknown, lot #: unknown, quantity: unknown).This report is for one (1) 5.5 exp verse fen scr 6.0x40.This is report 3 of 4 for complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
It was further reported that hardware removal procedure was completed successfully without any issues or difficulty.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: kwp, kwq.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Upon visual inspection of the returned device, the threads are torn inside the tulip head.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.This complaint is confirmed as the device threads are torn.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.
|
| |
|
Search Alerts/Recalls
|