Model Number 199723640
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional device product code: mni.(b)(4).The device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that the stabilization system implanted on the patient was dismantled due to the breaking of two tightening nuts and it was necessary to re-operate the patient to remove and replace the faulty system.The devices were removed.No further information provided.Upon receipt of the devices at manufacturing site it was identified that 5.5 exp verse fen screw has torn threads.Concomitant devices: unknown set screw (part #: unknown, lot #: unknown, quantity: unknown) and unknown rod (part #: unknown, lot #: unknown, quantity: unknown).This report is for one (1) 5.5 exp verse fen scr 6.0x40.This is report 3 of 4 for complaint (b)(4).
 
Event or Problem Description
It was further reported that hardware removal procedure was completed successfully without any issues or difficulty.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional device product codes: kwp, kwq.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Upon visual inspection of the returned device, the threads are torn inside the tulip head.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.This complaint is confirmed as the device threads are torn.No definitive root cause could be determined based on the provided information.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE FEN SCR 6.0X40
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9565134
Report Number1526439-2020-00279
Device Sequence Number19457950
Product Code NKB
UDI-Device Identifier10705034403090
UDI-Public(01)10705034403090
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K173095
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Model Number199723640
Device Catalogue Number199723640S
Device Lot Number238176
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Initial Date Received by Manufacturer 12/30/2019
Supplement Date Received by Manufacturer01/28/2020
01/31/2020
Initial Report FDA Received Date01/08/2020
Supplement Report FDA Received Date01/29/2020
02/03/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN LOCKING/SET SCREWS; UNKNOWN RODS
Outcome Attributed to Adverse Event Required Intervention;
-
-