| Model Number |
7205519 |
| Medical Device Problem Code |
Failure to Align (2522)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Foreign zipcode (b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that during an acl surgery, the acufex drill guide was not lined up.There was not a back-up available and it is unknown how the procedure was completed.There was no delay and no patient injuries reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
H10.H3, h6: the reported director drill guide aimer, used in treatment, has not been returned for evaluation.Without the reported product a visual or functional evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the aimer did not align properly.An exact root cause cannot be determined with confidence without the return of the device.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the manufacturing and complaint records was performed for the reported lot, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|