• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. LIPASE REAGENTS (4 X 160 TESTS); ADVIA CHEMISTRY XPT - LIPASE (LIP)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. LIPASE REAGENTS (4 X 160 TESTS); ADVIA CHEMISTRY XPT - LIPASE (LIP) Back to Search Results
Model Number ADVIA CHEMISTRY XPT - LIPASE (LIP)
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens healthcare diagnostics, inc.Has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using advia chemistry lipase reagent lot 485700.Preliminary investigation has indicated that not all cartons within this reagent lot are impacted.An urgent medical device recall (umdr) chc 20-03.A.Us was sent to us customers and an urgent field safety notice (ufsn) chc 20-03.A.Ous was sent to ous customers in december 2019.The umdr and ufsn advise customers of the investigation with the advia chemistry lipase reagent lot 485700 and provide instructions to help determine if the lipase reagent cartons in their inventory are impacted.If the cartons in the customers inventory are not impacted, the customers are instructed to proceed to using the cartons according to the instructions for use.If the cartons in the customers inventory are impacted, the customers are instructed to discard the impacted cartons and can request replacement.The next reagent lot is expected to be available by february 2020.If replacement kits are not available, customers are instructed to test patient samples using an alternate methodology.
 
Event or Problem Description
The customer obtained imprecise lipase (lip) patient and quality control (qc) results while using advia chemistry xpt and the customer also saw high imprecision for rbl value of calibration with lot #485700.The initial and repeat patient results were not reported to the physician(s).The customer used patient serum samples that were already resulted with previous lip lots to assess the precision of lip lot 485700 on advia chemistry.The customer also performed a lot-to-lot, instrument-to-instrument comparison with both qc and patient samples to investigate the observed imprecision.There are no known reports of patient intervention or adverse health consequences due to the discordant lip results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIPASE REAGENTS (4 X 160 TESTS)
Common Device Name
ADVIA CHEMISTRY XPT - LIPASE (LIP)
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
MDR Report Key9566534
Report Number2432235-2020-00012
Device Sequence Number16716105
Product Code CHI
UDI-Device Identifier00630414482446
UDI-Public00630414482446
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 01/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2020
Device Model NumberADVIA CHEMISTRY XPT - LIPASE (LIP)
Device Catalogue Number10311896
Device Lot Number485700
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2019
Initial Report FDA Received Date01/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2432235-01/02/2020-001-R
Patient Sequence Number1
-
-