| Model Number |
WRIST DRILL GUIDE |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/10/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopic tfcc repair the arthrex devices ar-8816g-02 an ar-8816g were used.The sleeve does not lock into the target device.It's slipping and it's unstable.The patient was already under anesthesia.The surgeon had to cancel the surgery as he had no backup equipment.The second surgery took place on the 17th of december.No further information received.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint confirmed, even though the slot was found to meet dimensional specifications, the cam was found to be worn and the device did not lock with the returned ar-8816g-02.A likely cause of the event is wear of the device.
|
| |
|