• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC DRIVING CAP/THREADED; MISC ORTHO SURGICAL INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC DRIVING CAP/THREADED; MISC ORTHO SURGICAL INSTR Back to Search Results
Model Number 03.010.486
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional narrative: the investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during an unknown procedure on (b)(6) 2019, the insertion handle and the driving cap would not thread together.The two reported devices were able to make work by holding the top of each devices.It is unknown if there was a surgical delay.Patient status and surgical outcome are unknown.This report is for one (1) driving cap/threaded this is report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation flow: functional/device interaction and damage.Visual inspection: the driving cap/threaded (p/n: 03.010.523, lot number: 8953363) was received at us cq.Visual inspection of the complaint device showed that the device was unable to assemble with a mating device (03.010.486, pi-15768036385691829) due to the complaint device and the mating device having damaged threads.Device failure/defect identified? yes.Functional test: the complaint device cannot be assembled with its returned mating device.The complaint was able to be replicated.Dimensional inspection: no dimensional inspection can be performed due to post-manufacturing damage and the nature of the complaint condition.Document/specification review: no design issues or discrepancies were identified.Complaint confirmed? yes.Investigation conclusion: this complaint is confirmed as the complaint device cannot be assembled with the mating device and has damaged threads.No root cause could definitively be determined for the reported complaint condition.No new, unique or different patient harms were identified as a result of this evaluation.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot: part: 03.010.523.Lot: 8953363.Manufacturing site: bettlach.Release to warehouse date: july 29, 2014.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVING CAP/THREADED
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9569500
Report Number2939274-2020-00109
Device Sequence Number16238737
Product Code LXH
UDI-Device Identifier10886982069047
UDI-Public(01)10886982069047
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.010.486
Device Catalogue Number03.010.523
Device Lot Number8953363
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2020
Initial Date Received by Manufacturer 12/13/2019
Supplement Date Received by Manufacturer01/27/2020
Initial Report FDA Received Date01/09/2020
Supplement Report FDA Received Date02/03/2020
Patient Sequence Number1
-
-