| Model Number |
27290M |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No indication for a material or manufacturing related issue was found during the investigation.The root cause most likely is a combination of corrosion due to insufficient cleaning and high force application, which led to overloading of the instrument and breakage.
|
| |
|
Event or Problem Description
|
|
Per the factory in (b)(6), allegedly, there was an event occurred in (b)(6) during a procedure, a piece broke off and fell inside the patient; it is unclear if the broken piece remained in the patient.
|
| |
|