| Model Number |
FG-0612 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis; therefore a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2019 that a symphion controller was used in a procedure performed on (b)(6) 2019.According to the complainant, during the procedure and outside the patient, the resecting device was noted to be activating on it's own without the foot pedal being pressed.The procedure was cancelled due to this event.There were no serious injury nor adverse patient effects reported as a result of this event.
|
| |
|