• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ASCENDA; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE Back to Search Results
Model Number 8780
Device Problem Separation Failure (2547)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/07/2019
Event Type  malfunction  
Manufacturer Narrative
Analysis identified damage occurred to the guidewire during the implant procedure.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare provider (hcp) via a manufacturer representative on (b)(6) 2019 regarding a patient receiving unknown medication (dose and concentration unknown) via an implanted infusion pump.The indication for use was non-malignant pain.It was reported that during the implant procedure the catheter stylet clubs could not be removed from the catheter and the hcp was unable to implant.Another catheter was opened and used.There were no known environmental, external, or patient factors that may have led or contributed to the issue and there were no applicable diagnostics or troubleshooting performed.The issue was considered resolved at the time of report and the patient's status was alive - no injury.No further complications were reported or anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCENDA
Type of Device
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key9571074
MDR Text Key174889180
Report Number3004209178-2020-00688
Device Sequence Number1
Product Code LKK
UDI-Device Identifier00763000051112
UDI-Public00763000051112
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/29/2021
Device Model Number8780
Device Catalogue Number8780
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/02/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/29/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age68 YR
-
-