| Model Number |
3092-020 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330); Limited Mobility Of The Implanted Joint (2671); No Code Available (3191)
|
| Date of Event |
03/29/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised to address pain, discomfort and loosening of the tibial component at bone to cement and cement to implant interface.Unknown cement was used.Doi: 2-3 years ago.Dor: (b)(6) 2019, left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : device history reviewed 07 oct 19, 0 non-conformances on this lot number, final micro and sterility tests passed.(b)(4) units released, expiry date: 30 nov 17.
|
| |
|