Model Number HARH45
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to stryker sustainability solutions for evaluation.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: jaws/blade subassembly damage or separation, incidental and prolonged activation against solid surfaces, such as bone, metal or plastic.The instructions for use (ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
 
Event or Problem Description
It was reported during a laparoscopic cholecystectomy, the harmonic ace +7 shears blade had broken off and fell into the patient.No harm occurred and all pieces were retrieved with a grasper.A larger incision was not needed for retrieval.There was no patient injury or medical intervention reported.The complainant is not aware of any extended procedure time.These are commonly used devices that are readily available.
 
Additional Manufacturer Narrative
Updates made to the following sections: d4, d10, g4, g7, h2, h3, h4, h6, h10, h11 the complaint device was returned to stryker sustainability solutions after the initial mdr was filed.Visual inspection revealed evidence of clinical use.The distal end of the blade was fractured and separated from the rest of the device.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: - jaws/blade subassembly damage - incidental and prolonged activation against solid surfaces, such as bone or plastic the instructions for use (ifu) state: - use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.- do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.- avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.- note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.- note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
 
Event or Problem Description
It was reported during a laparoscopic cholecystectomy, the harmonic ace +7 shears blade had broken off and fell into the patient.No harm occurred and all pieces were retrieved with a grasper.A larger incision was not needed for retrieval.There was no patient injury or medical intervention reported.The complainant is not aware of any extended procedure time.These are commonly used devices that are readily available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
SCALPEL, ULTRASONIC, REPROCESSED
Manufacturer (Section D)
STRYKER SUSTAINABILITY SOLUTIONS LAKELAND
5300 region ct
lakeland FL 33815
MDR Report Key9571914
Report Number0001056128-2020-00004
Device Sequence Number438272
Product Code NLQ
UDI-Device Identifier07613327346244
UDI-Public07613327346244
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K161693
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2022
Device Model NumberHARH45
Device Catalogue NumberHARH45RR
Device Lot Number10442402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2020
Initial Date Received by Manufacturer 12/12/2019
Supplement Date Received by Manufacturer01/22/2020
Initial Report FDA Received Date01/09/2020
Supplement Report FDA Received Date02/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
-
-