| Model Number |
HARH45 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned to stryker sustainability solutions for evaluation.As the device was not returned for evaluation, inspection was unable to be performed.A review of the dhr for the reported lot number supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: jaws/blade subassembly damage or separation, incidental and prolonged activation against solid surfaces, such as bone, metal or plastic.The instructions for use (ifu) state: use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.Do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.Avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.Note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.Note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.Should the device become available for return, the investigation will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported during a laparoscopic cholecystectomy, the harmonic ace +7 shears blade had broken off and fell into the patient.No harm occurred and all pieces were retrieved with a grasper.A larger incision was not needed for retrieval.There was no patient injury or medical intervention reported.The complainant is not aware of any extended procedure time.These are commonly used devices that are readily available.
|
| |
|
Additional Manufacturer Narrative
|
|
Updates made to the following sections: d4, d10, g4, g7, h2, h3, h4, h6, h10, h11 the complaint device was returned to stryker sustainability solutions after the initial mdr was filed.Visual inspection revealed evidence of clinical use.The distal end of the blade was fractured and separated from the rest of the device.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.Therefore, the most likely root causes are: - jaws/blade subassembly damage - incidental and prolonged activation against solid surfaces, such as bone or plastic the instructions for use (ifu) state: - use the torque wrench (already mounted to the shaft) to tighten the blade onto the hand piece.Turn the torque wrench clockwise while holding only the gray hand piece until it clicks twice indicating that sufficient torque has been applied to secure the blade.- do not attempt to bend, sharpen, or otherwise alter the shape of the blade.Doing so may cause blade failure and user or patient injury.- avoid contact with any and all other instruments while the instrument is activated.Contact with staples, clips or other instruments while the instrument is activated may result in cracked or broken blades.- note: do not use any other means than the torque wrench to attach or detach the instrument from the hand piece.- note: do not torque the instrument by hand without the torque wrench or damage may occur to the hand piece.The reported event will continue to be monitored through post-market surveillance.
|
| |
|
Event or Problem Description
|
|
It was reported during a laparoscopic cholecystectomy, the harmonic ace +7 shears blade had broken off and fell into the patient.No harm occurred and all pieces were retrieved with a grasper.A larger incision was not needed for retrieval.There was no patient injury or medical intervention reported.The complainant is not aware of any extended procedure time.These are commonly used devices that are readily available.
|
| |
|