MAUDE Adverse Event Report: COLOPLAST A/S TITAN TOUCH NB SCROT ZERO 18CM; INFLATABLE PENILE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COLOPLAST A/S TITAN TOUCH NB SCROT ZERO 18CM; INFLATABLE PENILE PROSTHESIS Back to Search Results
Model Number EN29181022
Device Problems Leak/Splash (1354); Material Split, Cut or Torn (4008)
Patient Problem No Information (3190)
Event Date 11/07/2019
Event Type  Injury  
Manufacturer Narrative
The lot number was reviewed for complaint trend, nonconforming reports and capa review.No trends were noted.Without the benefit of examination and testing, coloplast is precluded from commenting on the condition of the device or the cause of the occurrence.Should additional facts prompt us to alter or supplement any information or conclusions contained in the original mdr or in any prior supplemental reports, a follow-up report will be submitted.
 
Event Description
According to the available information, a pump tubing leak (tear) was noted.The patient had a revision surgery.
 
Manufacturer Narrative
This follow-up was created to document the additional event information.The device was not received for evaluation.Without the benefit of analyzing the device quality cannot confirm any observations or comment on the condition of the product.If the device become available, or additional information is received, the complaint will be re-evaluated according to procedures.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database revealed no trends for this lot.Review of nonconforming reports revealed no nonconformances for this lot.No capas are associated with this lot.
 
Event Description
Additional information received indicated that a leak from the tear was observed on the reservoir tubing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TITAN TOUCH NB SCROT ZERO 18CM
Type of Device
INFLATABLE PENILE PROSTHESIS
Manufacturer (Section D)
COLOPLAST A/S
holtedam 1
humlebaek, 3050
DA  3050
MDR Report Key9571972
MDR Text Key174475910
Report Number2125050-2020-00030
Device Sequence Number1
Product Code FHW
Combination Product (y/n)N
PMA/PMN Number
P000006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 02/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberEN29181022
Device Catalogue NumberEN2918
Device Lot Number6240167
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/11/2019
Initial Date FDA Received01/09/2020
Supplement Dates Manufacturer Received12/11/2019
Supplement Dates FDA Received02/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age57 YR
-
-