• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR TREK RX CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR TREK RX CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012276-20
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528); Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 12/17/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number e2019001.The balloon remains in the patient.The catheter is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The xience sierra stent referenced is being filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a calcified lesion in the tortuous distal right coronary artery (rca).The 3.50x38 mm xience sierra stent delivery system (sds) was being attempted to be placed distally; however, it could not cross the mid part of the vessel due to calcification.The physician wanted to remove the sds to pre-dilate the distal part of the vessel so the sds was removed.It was not noted that the stent had dislodged until pre-dilatation was being performed with the 3.50x20 mm trek balloon dilatation catheter (bdc) in the distal vessel and it was seen on angiography.Negative pressure was held for 2 seconds; however, the bdc would not completely deflate and became stuck in the same area as the stent in the calcification.Additionally, the balloon separated from the catheter.Multiple wire attempts were performed to try to remove the stent and balloon but were unsuccessful.Both the stent and balloon remain stuck in calcification in the vessel.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue; however, the reported difficulty removing the device, shaft separation, foreign body in patient and additional treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK RX CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9572176
Report Number2024168-2020-00528
Device Sequence Number1960600
Product Code LOX
UDI-Device Identifier08717648138584
UDI-Public08717648138584
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K103110
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number1012276-20
Device Catalogue Number1012276-20
Device Lot Number91002G1
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/17/2019
Supplement Date Received by Manufacturer02/29/2020
Initial Report FDA Received Date01/09/2020
Supplement Report FDA Received Date03/24/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
3.5X38 MM XIENCE SIERRA STENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age83 YR
-
-