| Model Number |
1012276-20 |
| Medical Device Problem Codes |
Deflation Problem (1149); Difficult to Remove (1528); Material Separation (1562)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
12/17/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2019001.The balloon remains in the patient.The catheter is not returning.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The xience sierra stent referenced is being filed under a separate medwatch report number.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a calcified lesion in the tortuous distal right coronary artery (rca).The 3.50x38 mm xience sierra stent delivery system (sds) was being attempted to be placed distally; however, it could not cross the mid part of the vessel due to calcification.The physician wanted to remove the sds to pre-dilate the distal part of the vessel so the sds was removed.It was not noted that the stent had dislodged until pre-dilatation was being performed with the 3.50x20 mm trek balloon dilatation catheter (bdc) in the distal vessel and it was seen on angiography.Negative pressure was held for 2 seconds; however, the bdc would not completely deflate and became stuck in the same area as the stent in the calcification.Additionally, the balloon separated from the catheter.Multiple wire attempts were performed to try to remove the stent and balloon but were unsuccessful.Both the stent and balloon remain stuck in calcification in the vessel.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue; however, the reported difficulty removing the device, shaft separation, foreign body in patient and additional treatment appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|