• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. BIOSURE SYNC 7-8MM TIBIAL FIXATION; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. BIOSURE SYNC 7-8MM TIBIAL FIXATION; SCREW, FIXATION, BONE Back to Search Results
Model Number 72202745
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/02/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an acl reconstruction surgery the device was found fractured and it was removed with tweezers.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported 7-8 mm biosure sync tibial fixation device, used in treatment, has been returned for evaluation.Visual assessment of the device shows one of the four fins has broken off and was not returned for examination.The remaining fins are bent and twisted and are missing small pieces.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive force/torque applied to the device during use.Improper site preparation.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the manufacturing and complaint records was performed for the reported lot, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOSURE SYNC 7-8MM TIBIAL FIXATION
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key9572199
Report Number1219602-2020-00060
Device Sequence Number2082100
Product Code HWC
UDI-Device Identifier03596010648983
UDI-Public03596010648983
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093943
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/21/2023
Device Model Number72202745
Device Catalogue Number72202745
Device Lot Number50715881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Initial Date Received by Manufacturer 12/20/2019
Supplement Date Received by Manufacturer02/19/2020
Initial Report FDA Received Date01/09/2020
Supplement Report FDA Received Date02/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age26 YR
-
-