| Model Number |
72202745 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/02/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an acl reconstruction surgery the device was found fractured and it was removed with tweezers.The procedure was successfully completed without delay using a back-up device.No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported 7-8 mm biosure sync tibial fixation device, used in treatment, has been returned for evaluation.Visual assessment of the device shows one of the four fins has broken off and was not returned for examination.The remaining fins are bent and twisted and are missing small pieces.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: excessive force/torque applied to the device during use.Improper site preparation.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the manufacturing and complaint records was performed for the reported lot, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|
| |
|