MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT; BALLOON EXPANABLE VASCLUAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT; BALLOON EXPANABLE VASCLUAR COVERED STENT Back to Search Results
Model Number LSMU0800637
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was not provided, a manufacturing review could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.
 
Event or Problem Description
It was reported that during treatment of the iliac via groin access, the delivery system allegedly could not be inserted through the introducer sheath after a few attempts.It was further reported that another device was used to complete the procedure.There was no reported patient injury.
 
Event or Problem Description
It was reported that during treatment of the iliac via groin access, the delivery system allegedly could not be inserted through two different introducer sheaths after a few attempts.It was further reported that another device was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not performed as the lot number was not provided.Investigation summary: the investigation is unconfirmed for the reported device incompatibility issue.Although there was difficulty encountered, the device successfully passed through an own 6fr terumo radifocus sheath during the evaluation.The complainant sheaths and stent sleeve were not returned.A 0.035" guidewire was also successfully inserted through the device during evaluation, though there was some resistance as it passed through the kink on the inner close to the proximal marker band.The presence of dried blood on the guidewire after it exited the distal tip likley also contributed to the resistance.The root cause could not be determined based upon the available information received from the field communications and the sample evaluation.It is unknown if there were procedural or handling techniques that contributed to the reported event.Labeling review: the ifu for the lifestream product was reviewed and contains the following information relevant to the reported event: precautions ¿ the device should only be used by physicians who are trained in endovascular procedures and are familiar with the complications, side effects and hazards of peripheral vascular interventions.¿ prior to device use, refer to the covered stent sizing table on the label and read the instructions for use.¿ crossing the implant with catheters or other adjunct devices can result in covered stent dislodgement or damage.Potential patient/device adverse effects that may occur include, but are not limited to, the following: ¿ covered stent dislodgement from balloon during tracking procedure ¿ covered stent misplacement during placement procedure directions for use: site access and preparation ¿ using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.Covered stent size selection ¿ select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation ¿ carefully remove the selected device from the package.¿ inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.¿ flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system air evacuation ¿ a 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.¿ with the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.¿ induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.¿ attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.¿ verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent ¿ advance the endovascular system over the guidewire into the introducer sheath.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT
Common Device Name
BALLOON EXPANABLE VASCLUAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
MDR Report Key9572355
Report Number9616666-2019-00155
Device Sequence Number19680321
Product Code PRL
UDI-Device Identifier05391522081024
UDI-Public(01)05391522081024
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P160024
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLSMU0800637
Device Catalogue NumberLSMU0800637
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/13/2019
Supplement Date Received by Manufacturer02/20/2020
Initial Report FDA Received Date01/09/2020
Supplement Report FDA Received Date02/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
Patient Weight60
-
-