• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number FL-410320
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Failure of Implant (1924)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Patient reported left side rupture and device "upside down (turned around)." device remains implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STYLE 410 COHESIVE SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
CS 
MDR Report Key9575054
Report Number9617229-2020-00677
Device Sequence Number11407643
Product Code FTR
UDI-Device Identifier10888628002784
UDI-Public10888628002784
Combination Product (Y/N)N
PMA/510(K) Number
P040046
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/13/2020
Device Catalogue NumberFL-410320
Device Lot Number2854517
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/17/2019
Initial Report FDA Received Date01/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-