• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. LIPASE REAGENTS (4 X 160 TESTS); ADVIA 1800 CHEMISTRY - LIPASE (LIP)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. LIPASE REAGENTS (4 X 160 TESTS); ADVIA 1800 CHEMISTRY - LIPASE (LIP) Back to Search Results
Model Number ADVIA 1800 CHEMISTRY - LIPASE (LIP)
Medical Device Problem Code Imprecision (1307)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Siemens healthcare diagnostics, inc.Has received customer complaints regarding failed calibrations and increased imprecision of quality control and patient samples when using advia chemistry lipase reagent lot 485700.Preliminary investigation has indicated that not all cartons within this reagent lot are impacted.An urgent medical device recall (umdr) chc 20-03.A.Us was sent to us customers and an urgent field safety notice (ufsn) chc 20-03.A.Ous was sent to ous customers in december 2019.The umdr and ufsn advise customers of the investigation with the advia chemistry lipase reagent lot 485700 and provide instructions to help determine if the lipase reagent cartons in their inventory are impacted.If the cartons in the customers inventory are not impacted, the customers are instructed to proceed to using the cartons according to the instructions for use.If the cartons in the customers inventory are impacted, the customers are instructed to discard the impacted cartons and can request replacement.The next reagent lot is expected to be available by february 2020.If replacement kits are not available, customers are instructed to test patient samples using an alternate methodology.Mdr 2432235-2020-00018 was filed for the same event.
 
Event or Problem Description
The customer observed calibration failures due to reagent blank limit (rbl) failures and imprecise quality control (qc) results while using lipase (lip) lot #485700 with advia 1800 chemistry system.The customer performed calibration with two different calibrator lots which resulted in the same imprecise qc.The customer did not provide any patient results.There are no known reports of patient intervention or adverse health consequences due to the issue with lip reagent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIPASE REAGENTS (4 X 160 TESTS)
Common Device Name
ADVIA 1800 CHEMISTRY - LIPASE (LIP)
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
MDR Report Key9575404
Report Number2432235-2020-00033
Device Sequence Number2224221
Product Code CHI
UDI-Device Identifier00630414482446
UDI-Public00630414482446
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/28/2020
Device Model NumberADVIA 1800 CHEMISTRY - LIPASE (LIP)
Device Catalogue Number10311896
Device Lot Number485700
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2019
Initial Report FDA Received Date01/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
2432235-01/02/2020-001-R
Patient Sequence Number1
-
-