Model Number 3662
Device Problem Difficult to Insert (1316)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/12/2019
Event Type  malfunction  
Manufacturer Narrative
Failure event observed during analysis.Final analysis found the root cause of the reported observation was due to the excess silicone adhesive that is preventing the lead to be fully inserted.
 
Event Description
It was reported that during the elective ipg upgrade the lead would not insert in the ipg due to some blockage.Failure event observed during analysis.Final analysis found the root cause of the reported observation was due to the excess silicone adhesive that is preventing the lead to be fully inserted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROCLAIM 7 ELITE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9575719
MDR Text Key174517325
Report Number3006705815-2020-00164
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067020222
UDI-Public05415067020222
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/06/2020
Device Model Number3662
Device Catalogue Number3662
Device Lot NumberA000071717
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/08/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Weight93
-
-