| Model Number |
6531 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: the device was returned for analysis.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.The device was removed from the guidezilla ii.Visual examination revealed that the balloon is separated 1cm from the tip.There are multiple kinks along the hypotube.Microscopic examination revealed that the guidewire lumen is stretched from the tip to 9cm from the tip.There is multiple buckling in the stretched area.The guidewire was attempted to be removed but it appears to be stuck because of the damages so it was left in place.Inspection of the remainder of the device presented no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 19dec2019.A 12mmx3.50mm nc quantum apex was returned with no reported issues.However, device analysis revealed balloon separation.
|
| |
|