| Model Number |
SPB14 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/01/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint (b)(4).Additional 510k numbers: k011245 and k002188.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant could not achieve primary stability.The implant was removed.
|
| |
|
Event or Problem Description
|
|
Additional information identified that another implant was able to be placed, making this case a same day replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received on (b)(6) 2019 identified that another implant was able to be placed, making this case a same day replacement.This event no longer meets reportability requirements.Therefore, no further medwatch reports will be submitted for this event.
|
| |
|