SMITH & NEPHEW, INC. GII ART INS SPCR TRL 1-2 15MM; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71430332
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/06/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery the insert trial broke when impacting.No delay was reported and a backup device was available.
 
Additional Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.A small section fractured off the device and all pieces were returned.The device has numerous nicks and gouges in the surface.The device was manufactured in 2007 and shows signs of extensive wear / usage.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GII ART INS SPCR TRL 1-2 15MM
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9578275
Report Number1020279-2020-00197
Device Sequence Number16951179
Product Code LXH
UDI-Device Identifier03596010481283
UDI-Public03596010481283
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71430332
Device Catalogue Number71430332
Initial Date Received by Manufacturer 01/06/2020
Supplement Date Received by Manufacturer01/06/2020
Initial Report FDA Received Date01/10/2020
Supplement Report FDA Received Date02/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-