• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OES URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OES URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P5
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Physical Entrapment (2327); No Information (3190)
Date of Event 10/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported scope was not returned to the service enter for evaluation; therefore, the cause of the reported event cannot be determined a this time.As part of our investigation, multiple follow-ups were made to the user facility in an attempt to gather additional information on the reported event; however, no additional information was obtained.However, if additional information becomes available, this report will be supplemented accordingly.This reported event has been reported by the importer on mdr# 2951238-2020-00313.
 
Event or Problem Description
On december 19, 2019, the service center received medwatch report # (b)(4) that states, "patient was undergoing a cystoscopy, right retrograde pyelogram, right ureteroscopy, laser lithotripsy procedure.As the surgeon proceeded to remove the ureteral scope from the right ureter it became lodged in the mid to distal ureter.The surgeon could not free the scope without significant traction.Despite attempts, the scope remained impacted within the ureter.A 14 french foley was replaced into the bladder, and ureteral scope was secured to patient's right thigh.The patient was then transferred to tertiary facility for further evaluation and management.".
 
Additional Manufacturer Narrative
This report is being updated to provide additional information reported by the patient's attorney (reported in b5, b7), and corrected information (h6-health effect clinical code).
 
Event or Problem Description
Additional information received: the patient was hospitalized for a period of 4 days and the scope was able to be removed by the surgeon without surgical intervention.
 
Additional Manufacturer Narrative
This supplemental report was submitted to provide additional information from the legal manufacturer (lm) and to update the following sections: g3, g6, h2, h4, h6 and h10.The lm performed the device history records for the concerned device and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A review of the repair history indicated that there was no repair history within the past year.A review of the complaint history indicates there are no similar complaints and no similar failures from the user facility.The investigation was completed by the lm and determined that there is no manufacturing, material or processing related cause for this failure mode.The concerned scope was not returned for evaluation; therefore, the exact cause of the reported malfunction could not be conclusively determined.Olympus will continue to monitor complaints for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES URETERO-RENO FIBERSCOPE
Common Device Name
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key9578354
Report Number8010047-2020-00873
Device Sequence Number3042825
Product Code FGB
UDI-Device Identifier04953170286988
UDI-Public04953170286988
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K912120
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberURF-P5
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/19/2019
Supplement Date Received by Manufacturer07/09/2021
09/07/2021
Initial Report FDA Received Date01/10/2020
Supplement Report FDA Received Date08/03/2021
09/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age60 YR
Patient Weight79
-
-