MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAL
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 05/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump passed the rewind, basic occlusion, prime, compromised forced sensor system and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.(b)(4).
 
Event or Problem Description
The customer reported via phone that the insulin pump had motor error alarm.Customer¿s blood glucose was over 300 mg/dl at the time of incident.Customer was able to successfully clear the alarm.Customer stated that the insulin pump was able to rewind.Drive support cap was normal.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAL
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9578865
Report Number3004209178-2020-54105
Device Sequence Number2902545
Product Code OYC
UDI-Device Identifier00613994743916
UDI-Public(01)00613994743916
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAL
Device Catalogue NumberMMT-723NAL
Device Lot NumberA3723NALJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/09/2020
Initial Report FDA Received Date01/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age27 YR
Patient Weight123
-
-