MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAH; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-722NAH
Medical Device Problem Codes Mechanical Problem (1384); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/27/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Insulin pump passed the rewind, basic occlusion, prime/a33 and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.Pump received with cracked case display window corner.(b)(4).
 
Event or Problem Description
The customer reported via phone that the insulin pump had motor error alarm.Customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump was able to rewind.Drive support cap was normal.Customer stated that the top part going to screen was cracked.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP MMT-722NAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9579081
Report Number3004209178-2020-54305
Device Sequence Number14178496
Product Code OYC
UDI-Device Identifier00613994622587
UDI-Public(01)00613994622587
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-722NAH
Device Catalogue NumberMMT-722NAH
Device Lot NumberA2722NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/08/2020
Initial Report FDA Received Date01/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Patient Weight203
-
-