| Model Number |
MMT-722NAH |
| Medical Device Problem Codes |
Mechanical Problem (1384); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Insulin pump passed the rewind, basic occlusion, prime/a33 and displacement test.Pump received with stuck motor error alarm loop during bolus delivery.The motor was tested outside of the device and passed.The motor may have had and intermittent failure that was not detected during our testing.Pump received with cracked case display window corner.(b)(4).
|
| |
|
Event or Problem Description
|
|
The customer reported via phone that the insulin pump had motor error alarm.Customer¿s blood glucose was unknown at the time of incident.Customer stated that the insulin pump was able to rewind.Drive support cap was normal.Customer stated that the top part going to screen was cracked.The insulin pump will be returned for analysis.
|
| |
|