| Catalog Number |
6973060 |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent cervical disc arthroplasty at c6-7 due to degenerative disk disease(ddd).Intra op, inserter did not retain the implant properly when impacted with the mallet.There were no patient complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis results: the upper (superior) and lower (inferior) tips of the inserter shaft have been bent to where one side at the.4mm dimension located on the print at location f6 on page 1 of 2 appear to be touching.This is consistent with over-torque if information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Mdr decision corrected to not reportable.No additional supplemental reports will be required unless additional information received indicates reportable event.
|
| |
|
Additional Manufacturer Narrative
|
|
H1: review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunction.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|